#Product Trends
Mastering Clean Air with CLINICAIR® 1B
The CLINICAIR® 1B is our most compact air handling unit, specially designed to be installed directly inside the room or as close as possible to the risk area to be treated.
One of its possible configurations includes a return air grill on the front face, allowing installation without a return duct. The clean air is then supplied from the top : either via a plenum or a very short duct connected to a Laminar Air Flow ceiling, allowing you to reach ISO 5 air quality in the room.
Main Advantages of In-Room Installation
✅ Lower risk of airborne infection
The drastically reduced ducting length prevents particle retention and the development of microorganisms in the duct system.
✅ Lower installation costs
No technical room required and less ducting needed. It’s also ideal for rooms with challenging architecture (e.g. no false ceiling).
Key Features
- Airflow up to 2,500 m³/h
- 3-stage filtration including HEPA14 filter
- BIOXIGEN® technology for microorganisms’ removal
- Integrated PLC control for T°C, RH%, pressure, and more
- Compliance with: ISO 14644-1,NFS 90-351and the ASHRAE 170 standard
CLINICAIR® 1B is a dedicated device designed to treat and ensure air quality conditions within critical medical environments, including operating rooms.
For a customized approach of your project, contact our team !