video corpo
Add to favorites

#Trade Shows & Events

On-site Technical Support for Intraoperative Neuromonitoring System in Indonesia

Our team provided on-site technical support to a partner client in Indonesia who is utilizing our Cynapse IONM 32CH with EP system for intraoperative neuromonitoring.

While utilizing the Somatosensory Evoked Potential (SEP) module to collect intraoperative signals, we encountered two major issues that severely disrupted the surgical monitoring process.
1. Incomplete waveform collection: Electrical stimulation was applied to the patient’s feet and signals were recorded at the head. The total signal transmission duration was approximately 60 ms, which exceeded the system’s default 50 ms acquisition window, resulting in incomplete waveform capture.
2. Severe signal interference: Dense jagged noise appeared on the baseline. The peaks and troughs of waveforms could not be distinguished clearly, making the monitoring data clinically invalid.

Details

  • 555 Qing Dai Lu, Pu Dong Xin Qu, Shang Hai Shi, China, 201318
  • NCC Medical